Health Canada Medical devices

SmartDrive MX2+

Issued August 13, 2025
Issue
Max Mobility has initiated a voluntary correction of the speed control dial, a component utilized with the SmartDrive MX2+ Power Assist device (this attaches to parent device manual wheelchairs to provide power assist). Through market feedback and subsequent investigation, Max Mobility has identified that a material change with the printed circuit board assembly (PCBA) has caused performance issues and a second issue with cable connectivity with the speed control dial. In December 20, 2024 Permobil issued a recall the speed control dial potentiometer had a material changes in the potentiometer that resulted in voluntary recall due to allegations of continued drive and involuntary movement. (Health Canada Recall RA-76773). Recall start date: August 12, 2025
Hazard description
Performance
Product
Product name
SmartDrive MX2+
Reference
ID
RA-77894