Health Canada Medical devices

Merit basixCOMPAK™, PhD™ Hemostasis Valve, and basixSKY™ Inflation Syringe

Issued August 14, 2025
Issue
Merit Medical became aware of a manufacturing defect after an increased number of complaints. Investigation into the issue concluded that the lack of fluorinated silicone in the bleed back seal (the secondary valve) caused by the mix of incorrect silicone during production line influenced the performance of the PhD™ Hemostasis Valve. The lack of silicone was found to prevent the secondary valve from releasing and closing properly. The valve cap potentially would remain depressed causing leakage. Recall start date: July 30, 2025
Hazard description
Performance
Product
Product name
Merit basixCOMPAK™, PhD™ Hemostasis Valve, and basixSKY™ Inflation Syringe
Reference
ID
RA-77904