Beckman Coulter has determined that on the DxC 500i clinical analyzer, when ordering a clinical chemistry (CC) combination test simultaneously with an immunoassay (IA) test, if the IA test is processed between the CC constituent tests, the sample status of the CC test will remain "in progress" with no error events. Recall start date: August 4, 2025