Back to RecallTracker
Health Canada Medical devices

Total Knee Arthroplasty (TKA) and Total Hip Arthroplasty (THA)

Issued August 28, 2025
Issue
Stryker discovered specific to the applications listed above, an increase in the software error #3 (se3) error code when a Mako® system shutdown or a Mako® system restart is not performed prior to switching between applications (i.e. TKA to THA). Recall start date: August 18, 2025
Hazard description
Performance
Product
Product name
Total Knee Arthroplasty (TKA) and Total Hip Arthroplasty (THA)
Reference
ID
RA-77996