Health Canada Medical devices

Getinge Washer/Disinfector

Issued August 28, 2025
Issue
The issue found can cause the device to overheat during operation if the circulation pressure calibration was not performed or been performed incorrect and the low-pressure alarm is not triggered. The scope of this action is the electrical heated 88-5 washer disinfectors. Justification: the construction on steam heated devices is different and the heating elements cannot be overheated. This field action is initiated to verify that the circulation pressure in standby mode reads 0.0 kPa. If the value deviates from 0.0 kPa, the device must be recalibrated by a qualified technician to ensure safe continued clinical use. Recall start date: August 21, 2025
Hazard description
Performance
Product
Product name
Getinge Washer/Disinfector
Reference
ID
RA-77997