MAMMOMAT Inspiration

Issued October 24, 2025
Issue
The operator table (part number 14467741) is not intended and not released for bus installations as it was designed for stationary operation only. Due to the impacts and/or vibrations during transportation, there is a risk that parts of the operator table (e.g. screws, monitor, monitor holder) could come loose and potentially fall off resulting in a risk of serious injury to the operating staff from falling parts. In addition, damage to the lead shield may also occur.
Hazard description
Device compatibility
Product
Product name
MAMMOMAT Inspiration
Reference
ID
RA-78385