There is a potential for delayed time to high-voltage (HV) therapy should a rare sequence of events occur. Through 2-October-2025, six (6) events (delay of 2-17 seconds) were observed among 4,900 implanted devices worldwide (~0.12%). Five of the six events experienced this behavior during a controlled defibrillation test. While not observed clinically, a delay in HV therapy may impact defibrillation efficacy.