Medtronic Restore™ Clinician Programmer Application

Issued November 28, 2025
Issue
Medtronic is initiating a medical device recall (notification) to notify consignees of a potential for a software issue that may occur when attempting to clear a device reset message using the model A71100 Restore Clinician Programmer application (CP App). The A71100 CP App is used to program the restore family of Spinal Cord Stimulation implantable neurostimulators (INS). A device reset can occur for a variety of reasons, including when the INS has an overdischarged battery, which is normally cleared during a programming session. In rare cases, the device reset message cannot be cleared and may result in the inability to resume therapy resulting in the patient experiencing a recurrence of underlying pain symptoms. Resolution would require surgical replacement of the INS to resume therapy.
Hazard description
Performance
Product
Product name
Medtronic Restore™ Clinician Programmer Application