The action is being initiated due as a cybersecurity vulnerability was discovered as part of an internal SW penetration test. The USB interface was identified as a possible gateway. There have not been any reports of patient injuries to date.
Hazard description
Performance
Product
Product name
Zeus Infinity Empowered (Anesthesia Workstation)
Reference
ID
RA-81338