Beckman Coulter has confirmed an error in SW version 2.3.90.0 that causes samples to be dispensed unexpectedly close to the bottom of the output tubes. The intended design is for samples to be dispensed above the fluid surface into the output tubes.
Hazard description
Performance
Product
Product name
CellMek SPS
Reference