Health Canada Medical devices

Alphenix Systems

Issued February 10, 2026
Issue
An internal error would occur if F-STORE or image acquisition is performed while CAAS clinical analysis function is used. The manufacturer confirmed that this is caused by a software problem and will not occur if F-STORE or image acquisition are performed after terminating CAAS clinical analysis function.
Hazard description
Performance
Product
Product name
Alphenix Systems
Reference
ID
RA-81590