Back to RecallTracker
Health Canada Medical devices

Sterile Contro-Vac Whistle Tip Suction Cath W/2 Gloves & Pop-Up Cup Vinyl

Issued February 27, 2026
Issue
Medline has identified that a smaller open Suction Catheter (8Fr), may have been packed within the 14Fr Suction Catheter Kits. Use of a smaller size suction catheter may lead to difficulty or inability to aspirate liquids or semisolids from a patient's upper airway and may result in hypoxia and/or respiratory failure. Customers are advised to discard this product.
Hazard description
Device compatibility - Labelling and packaging
Product
Product name
Sterile Contro-Vac Whistle Tip Suction Cath W/2 Gloves & Pop-Up Cup Vinyl
Reference
ID
RA-81661