AneurysmFlow

Issued April 2, 2026
Issue
Philips has determined that the mafa ratio does not provide reliable prognostic information regarding aneurysm occlusion following flow diverter stent treatment. the instructions for use (IFU) of AneurysmFlow state that the mafa ratio should not be used for making clinical decisions (section 9.2 measuring flow). In addition, when hovering over the mafa ratio header within the software interface, an on-screen message states that the mafa ratio is not to be used for clinical decision making. despite the existing cautions, the displayed mafa ratio may still be considered during intraprocedural decision-making and could influence clinical judgment. This may potentially result in an incorrect clinical decision.
Hazard description
Performance
Product
Product name
AneurysmFlow
Reference
ID
RA-81836