Mixing Bowl And Spatula, 20- Pack

Issued April 2, 2026
Issue
Product Mixing Bowl And Spatula, 20- Pack Issue Medical devices - Sterility What to do Contact the manufacturer if you require additional information. Audience Healthcare
Hazard description
Zimmer Biomet is conducting a lot specific medical device recall removal for the mixing bowl and spatula. Ten complaints have been received identifying issues at the time of use related to the package seal, including incomplete seals, wrinkles in the seals or peeling seals. Seal integrity testing conducted internally showed potential sterility breach in 7.5% of the tested units that had visual seal nonconformities. The devices are distributed in 20-pack boxes. The defectiveness was discovered on November 19, 2025. Issue was identified through the complaints handling process.
Product
Product name
Mixing Bowl And Spatula, 20- Pack
Reference
ID
RA-81888