Health Canada Medical devices

SynchroMed Clinician Programmer Application

Issued April 8, 2026
Issue
Product SynchroMed Clinician Programmer Application Issue Medical devices - Performance What to do Contact the manufacturer if you require additional information.
Hazard description
Complaint investigations identified a software sequencing issue in the Medtronic A810 SynchroMed Clinician Programmer Application (CP app), specifically affecting flex dosing when programming the SynchroMed 8637 and 8667 pumps. When flex dosing steps are added overlapping and then out of chronological order, the app may fail to reorder them correctly, resulting in infusion steps being delivered at incorrect times. This can lead to missed flex doses, underdosing, or shifts in dose timing, and no immediate alerts. The issue is limited to flex dosing and does not impact other infusion modes. Eleven (11) complaints have been reported globally related to this issue.
Product
Product name
SynchroMed Clinician Programmer Application