Edwards Lifesciences Swan-Ganz Catheters

Issued April 8, 2026
Issue
Product Edwards Lifesciences Swan-Ganz Catheters Issue Medical devices - Performance What to do Contact the manufacturer if you require additional information. Audience Healthcare
Hazard description
There have been a growing number of complaints (outside of Canada) of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on Swan-Ganz catheters. The issue is linked to manufacturing process and material changes and is associated with situations where the (blue) proximal injectate hub and lumen experience frequent manipulation or flexing. Most complaints are associated with performance of bolus thermodilution cardiac output procedures. This has led to instances of leakage, lumen damage and breakage and may lead to infection, medication loss, and/or blood loss.