Back to RecallTracker
Health Canada Health products

Vivo 45 LS

Issued April 17, 2026
Issue
Product Vivo 45 LS Issue Medical devices - Performance What to do Contact the manufacturer if you require additional information. Audience Healthcare
Hazard description
In devices with the affected firmware versions, treatment in volume-controlled ventilation modes (VCV, VCV-SIMV and VCV-MPV) can, under specific trigger conditions, result in delivered tidal volume being reduced significantly below the set value.
Product
Product name
Vivo 45 LS
Reference
ID
RA-82020