Back to RecallTracker
Health Canada Health products

Visian® ICL™/TICL™ Online Calculation Software

Issued April 21, 2026
Issue
Product Visian® ICL™/TICL™ Online Calculation Software Issue Medical devices - Performance What to do Contact the manufacturer if you require additional information. Contact your STAAR Sales Representative or Customer Service at customerservice.ag@staar.com or by phone at +41323328888 Monday through Friday, 8:00am to 5:00pm CET. Audience Healthcare
Hazard description
Between April 13 and 16, 2026, the Visian® ICL™/TICL™ online calculation software may have generated incorrect Implant Orientation Diagram (IOD) printouts used for EVO/EVO+ Toric ICL implantation. The issue affected only the printed diagram and not the lenses themselves.
Product
Product name
Visian® ICL™/TICL™ Online Calculation Software
Reference
ID
RA-82024