MOSAIQ

Issued April 28, 2026
Issue
Product MOSAIQ Issue Medical devices - Performance What to do Contact the manufacturer if you require additional information. Audience Healthcare
Hazard description
When using Anatomical Position Monitoring (APM) during a Comprehensive Motion Management (CMM) unity treatment delivery, if the user performs a baseline shift (BLS) plan and an error occurs during the plan transfer, in some instances, the BLS plan may not load into mosaiq resulting in the original beams being delivered without the new BLS applied.
What to do
Contact the manufacturer if you require additional information. Email: application.support.na@elekta.com Tel: +1 855 6935358 https://www.elekta.com