Health Canada Medical devices

Azurion Systems

Issued April 30, 2026
Issue
Product Azurion Systems Issue Medical devices - Performance What to do Contact the manufacturer if you require additional information. Audience Healthcare
Hazard description
Philips has identified that the float tabletop control on the control module of Azurion systems may experience mechanical wear over time. The float tabletop control is located on the control module at the table-side. When the control is actuated, the table brakes disengage, allowing adjustment of the longitudinal or transverse position of the table. Over time, mechanical wear may lead to detachment, cracking, or breakage of the float tabletop control, which may lead to reduced availability or unavailability of longitudinal and transverse tabletop movement, as well as the formation of sharp edges.
Product
Product name
Azurion Systems
Reference
ID
RA-82045