Exoplan

Issued April 28, 2026
Issue
Product Exoplan Issue Medical devices - Performance What to do DO NOT USE AFFECTED LIBRARY AS PER THE FIELD SAFETY NOTICE. Audience Healthcare
Hazard description
During internal testing of libraries an error was identified within specific Straumann surgical protocol libraries used with Exoplan version 3.1. Under certain implant/sleeve/workflow selections, the protocol output displays an incorrect instrument combination in the final drilling steps. In two consecutive protocol steps an incorrect drill key is displayed, which may result in a drilling depth up to 2 mm deeper than intended. This may lead to unintended injury depending on anatomical conditions and clinical use. The deviation is limited to specific implant/sleeve/workflow selections and does not affect all protocol configurations.
Product
Product name
Exoplan
Reference
ID
RA-82095