IceMan CLASSIC3™

Issued April 30, 2026
Issue
Product IceMan CLASSIC3™ Issue Medical devices - Performance What to do Customers are notified and provided instructions on follow up. Quarantine and return affected products. Audience Industry
Hazard description
Complaints received reporting a consistent failure mode described as a broken connector on IceMan CLASSIC™ cube unit. If the connector is broken, the unit may not function as intended, which could result in a temporary interruption of prescribed cold therapy until a replacement unit is provided. Leakage could saturate the barrier (dressings) and contact the surgical site with non-sterile water, which could in potential cold-related tissue injury or surgical site infection.