Hazard description
Philips has identified a potential safety issue affecting Allura R8.2 systems equipped with a AD7x patient table type potentially leading to loss of geometry movements. In some cases, following a system (re)start, manual and motorized geometry movements may be disabled due to a firmware issue in the Advanced Motion Controller (AMC) single-axis motion control unit used in the AD7x tables of Allura R8.2 systems. When the issue occurs, no message is displayed to the user. In this situation, X-ray imaging is available. The issue can be resolved by performing a cold restart of the system or by restarting the geometry.