Back to RecallTracker
Health Canada Health products

SPECTRALIS® with Flex Module

Issued August 7, 2026
Issue
Product SPECTRALIS® with Flex Module Issue Medical devices - Performance What to do Contact the manufacturer if you require additional information. Audience Healthcare
Hazard description
There is a possibility that the castor wheels of the Flex stand are not properly attached. The wheels may come loose. As a result, the Flex stand may tip over. Users shall schedule a service appointment as soon as possible so that the castor fastenings can be inspected and corrected.
Reference
ID
RA-82505