CareIntellect Perinatal

Issued September 21, 2026
Issue
Product CareIntellect Perinatal Issue Medical devices - Performance What to do Contact the manufacturer if you require additional information. Audience Healthcare
Hazard description
GE Healthcare has become aware of an issue in CareIntellect Perinatal (CIP) that may result in loss of monitoring information and clearance of active patient monitoring alarms. When this issue occurs due to an interruption in system connectivity, monitoring data will be unavailable for up to approximately 20 seconds and currently active visual and audible alarm indicators will no longer be displayed or announced at the CIP user interface. Following resolution of system connectivity, alarm triggers will restart based on the configured thresholds. This issue could delay recognition of a maternal or fetal condition requiring clinical assessment or intervention.
Reference
ID
RA-82711