Back to RecallTracker
Health Canada Health products

FreeO2 Automated Oxygen Therapy Device Adult & Pediatric - Version HW 2.3

Issued September 30, 2026
Issue
Product FreeO2 Automated Oxygen Therapy Device Adult & Pediatric - Version HW 2.3 Issue Medical devices - Performance What to do Contact the manufacturer if you require additional information. Audience Healthcare
Hazard description
Under specific conditions following a deep battery discharge, error 29515 may occur during startup of a FreeO2 HW 2.3 device using software package 1.4.2 and prevent its use. The update to software package 1.4.3 corrects this issue and restores the expected startup behaviour.
Product
Product name
FreeO2 Automated Oxygen Therapy Device Adult & Pediatric - Version HW 2.3
Reference
ID
RA-82749